Comparison of Photon-electron and Photon Radiotherapy for Supraclavicular Lymph Nodes of Mastectomy Patients with Left-sided Breast Cancer
Journal of Cancer Prevention 2020;25:48-54
Published online March 30, 2020
© 2020 Korean Society of Cancer Prevention.

Mehran Mohseni1, Fatholah Mohaghegh2, Nasrin Robatmili1, Mohammad Reza Bayatiani2, Fatemeh Seif2, Nayyer Sadat Mostafavi3

1Department of Medical Physics, School of Paramedical Sciences, Kashan University of Medical Sciences, Kashan, 2Department of Medical Physics and Radiotherapy, Arak University of Medical Sciences and Khansari Hospital, Arak, 3Department of Medical Physics, School of Paramedical Sciences, Isfahan University of Medical Sciences and Khansari Hospital, Isfahan, Iran
Correspondence to: Nasrin Robatmili, E-mail: robatmilyn@yahoo.com, https://orcid.org/0000-0001-9276-3029
Received November 22, 2019; Revised December 19, 2019; Accepted December 20, 2019.
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract
The aim of radiotherapy is to deliver the highest possible radiation dose to the tumor and the lowest radiation to normal tissues surrounding the tumor. In the present study, lymph nodes of the supraclavicular region were treated using two therapeutic techniques, namely photon technique (PT) and combinatory photon-electron technique (CPET). We recruited 50 patients with local lymph node metastasis. The photon energies were 6-15 MV. Furthermore, the electron beam energy was 18 MeV in CPET. The study findings revealed that the mean delivered dose to target volume was 41.12 짹 2.98Gy for PT and 44.56 짹 1.90Gy for CPET. The percentage of the target volume irradiated to 90% of the prescribed dose (V90) was calculated as 74.61% 짹 9.30% and 82.06% 짹 9.70% for PT and CPET, respectively. The mean dose delivered to the heart and lungs was not significantly different between the two groups. Furthermore, the maximum doses delivered to the spinal cord were 12.55Gy in PT and 8.89Gy in CPET. The mean doses delivered to the thyroid gland were 39.26 and 34.89Gy in PT and CPET. According to the study results, the maximum doses delivered to the spinal cord, head of the humerus bone, and thyroid were reduced significantly as measured the CPET technique. In contrast, no significant difference was observed regarding the dose delivered to the heart and lung. The dose delivered to the supraclavicular region determined by the CPET was significantly augmented. Furthermore, the coverage of the tumor mass was optimized using the new method.
Keywords : Breast cancer, Treatment planning, Supraclavicular region dose, Thyroid dose
INTRODUCTION

Breast cancer has been reported as the second fatal malignancy after lung cancer among women worldwide. The studies performed during 1989-2012 indicated that the number of patients affected by breast cancer increased up to 36% [1]. The common therapeutic techniques for breast cancer include surgery, chemotherapy, radiotherapy, and hormone therapy. It is noteworthy that approximately 60% of all patients worldwide are treated by radiotherapy [2,3].

During the course of treating cancerous cells, surrounding healthy tissues also get exposed to the radiated beam. Consequently, it is of considerable importance to concentrate the dose on the target mass, so that sensitive and healthy organs, such as the heart (especially, in left-sided breast cancers), lungs, spinal cord, thyroid, and the other breast, would be minimally exposed.

As a result, a proper treatment plan is one of the most important key components of effective therapy of breast cancer. For this, the type and intensity of radiation, beam direction, field measurements, and anatomical condition of the patient should be considered [4]. Due to the possibility of changes in anatomical regions, different therapeutic techniques are considered for different disease stages, among which the techniques of common three-dimensional tangential fields and combinatory fields could be plausible options. In these techniques, electron fields, photons, or a combination of both fields are utilized [5-9].

The most common radiotherapy method for breast cancer uses two opposing medial and lateral tangential fields of photon along with the wedge in chest wall region and the single-field photon in the supraclavicular region [10]. In the common method, the chest wall region is half-beamed from both sides of the lungs and supraclavicular region. Next, the dose distribution is optimized using a wedge filter and different beam energies.

The current study aimed to evaluate a therapeutic method in which two tangential opposing fields along with the wedge are used for the chest wall region, which are similar to the common three-dimensional method. A posterioranterior (PA) photon field and an anterior-posterior (AP) electron field opposing each other were considered for the supraclavicular region. The PA photon and the AP electron fields were selected because of the substantially high and low penetration depths of photons and electrons, respectively.

In the present study, the absorbed doses of supraclavicular lymph nodes of planning treatment volume (PTV) and the normal tissues surrounding the PTV were assessed. Finally, a comparison was made between the results of the two therapeutic techniques including the photon technique (PT, the common therapeutic technique) and the combinatory photon-electron technique (CPET), which were applied for the lymph nodes of the supraclavicular region [10]. In the common method, the chest wall region is half-beamed from both sides of the lungs and supraclavicular region. Next, the dose distribution is optimized using a wedge filter and different beam energies.

MATERIALS AND METHODS

This study was approved by the ethics committee of the Kashan University of Medical Sciences, Iran (code number: IR.KAUMS.REC.1395.151). In this study, data of 50 mastectomy patients with left breast cancer and with local lymph node metastasis were used. The study was performed in the Radiation Oncology Department at Ayatollah Khansary Hospital (Arak, Iran) between 2016 and 2017. Informed consent was obtained from all individuals participating in this study.

Computer treatment planning

Prior to the computed tomography (CT)-simulation procedure, the patient was laid on the breast board in a suitable angle. The angle was chosen, so that the chest wall was located parallel to the horizon. Afterwards, a marker was considered as the point of coordinates (x, y, z) and the wires specifying the entrance of medial and lateral fields, as well as the lower and upper borders of the breast were located on the patient’s body by the doctor.

The ipsilateral hand of the patient was positioned above his/her head, and the head was rotated toward the contralateral breast. The entrance point of the medial tangent was regarded as the midline, and the entrance point of the lateral tangent was considered 2 cm from the point of breast tissue touch. The initial data were recorded in the planning system after CT-simulation and preparation of CT images. The radiation dose was planned to be 50 Gy in 25 sessions.

Patient sampling was performed using the normal distribution function and simple random method. The total number of samples used was 50.

All the patients underwent radiation therapy including single PT for supraclavicular region using two opposing tangential fields along with the wedge for the chest wall region. To plan CPET for chest wall region using the ISOgray Treatment Planning System, two opposing tangential fields along with the wedge similar to the common three-dimensional technique were used.

On the other hand, the field for the supraclavicular region was planned as two-sided opposing fields of PA photon and AP electron with 15 MeV and 18 MeV energies, respectively. This system can plan treatment by various methods, such as three-dimensional conformal radiation therapy.

The delivered dose and the dose volume histogram (DVH) of the covered volume were obtained for each healthy organ and tissue in addition to the tumoral tissue using the ISOgray software. Two samples of dose distribution resulting from the two therapeutic methods for the tumoral tissue are demonstrated in Figure 1. The hot and cold parts, as well as the DVH, were extracted and analyzed clinically and physically by comparing with the standard protocols.

Dosimetric parameters

Some parameters, including V95 (%) and V90 (%) (the percentage of the target tissue volume irradiated to 95 and 90% of the prescribed dose, respectively); V30 (%), V25 (%), and V5 (%) (the percentage of the heart volume receiving 30, 25, and 5 Gy dose, respectively); and V40 (%) and V20 (%) (the percentage of the ipsilateral lung volume receiving 40 and 20 Gy dose, respectively), needed to be evaluated for dose distribution and were assessed using DVH. The DVH for all exposed organs (the heart, lungs, head of the humerus bone, thyroid, and spinal cord) and PTV (supraclavicular and chest wall) after performing both therapeutic methods are indicated in Figure 2. Moreover, parameters of maximum, minimum, and mean dose delivered to all exposed organs and PTV were determined using the aforementioned software. A sample of the data extracted from the ISOgray software is presented in Figure 3.

Statistical analysis

All statistical analyses were performed by SPSS ver. 16 (SPSS Inc., Chicago, IL, USA). The one-way ANOVA test was applied to compare the results obtained by dosimetric parameters of the two therapeutic methods. P < 0.05 was considered significant.

RESULTS

Figure 1 illustrate the extracted isodose curves for tumoral tissue, and Figure 2 shows the DVH for all organs (both tumoral and healthy tissues) using both techniques in a patient. Furthermore, Table 1 and 2 demonstrate the calculated dose parameters for the supraclavicular lymph nodes and normal tissues in both methods. The parameters, including maximum dose, mean dose for target volume (supraclavicular) and exposed organs (thyroid, spinal cord, lungs, heart, and head of the humerus bone); V95 and V90 for the target volume (supracla-vicular region); V40 and V20 for the lungs; and V5, V25, and V30 for the heart, were calculated and compared between the two methods.

Doses delivered to organs at risk

Table 1 lists the maximum and mean doses delivered to the spinal cord, thyroid, head of the humerus bone, lungs, and heart for 50 patients with left-sided breast cancer. The results of both the treatment plans were compared (Fig. 4 and 5). Compared with PT, CPET resulted in a reduction in the maximum and mean doses delivered to the spinal cord, thyroid, and head of the humerus bone. The maximum and mean doses delivered to the lungs and heart were not significantly different.

As shown in Table 2, the mean V25 (%) values for the heart are 4.37% ± 8.36% and 4.41% ± 7.07% in CPET and PT, respectively. Moreover, V30 (%) and V5 (%) were 3.82% ± 7.92% and 32.07% ± 9.75% in CPET, respectively, and 3.71% ± 6.81% and 31.41% ± 9.76% in PT, respectively. The P-value was not significantly different between the two techniques. Furthermore, as shown in Table 2, the mean V40 (%) values for the lung were 9.43% ± 7.81% in CPET and 10.29% ± 10.10% in PT, whereas V20 (%) values were 25.23% ± 10.78% and 26.98 ± 10.52% in CPET and PT, respectively.

Doses for planning treatment volume (supraclavicular nodal)

The mean of the maximum dose delivered to the supraclavicular region and the P-values for the differences of this parameter between the two techniques are shown in Table 1. It should be noted that there was no significant difference between the two methods regarding the maximum doses for the supra-clavicular region (P = 0.086). Moreover, the mean doses delivered to the supraclavicular region were 44.56 ± 1.90 and 41.12 ± 2.98 Gy for CPET and PT, respectively. According to the results of one-way ANOVA, the latter parameter (the mean doses) was significantly different between the two groups (P = 0.007). The maximum and mean doses delivered to the supraclavicular region increased in CPET compared with those in PT (Fig. 4 and 5). In addition, the dose delivered to the supraclavicular region was augmented (Fig. 6).

Based on Table 2 and Figure 6, the mean V95 (%) values were 65.80% ± 13.10% and 55.53% ± 12.31% for CPET and PT, respectively. Furthermore, as mentioned in Table 1, the two groups were significantly different with respect to V95 (%) (P = 0.002). Likewise, the mean V90 (%) values for the two methods are shown in Table 2 and Figure 6, and the difference between the two techniques with respect to this parameter was statistically significant (P = 0.031).

DISCUSSION

Electron beam therapy (EBT) is currently one of the commonly used radiotherapy methods. EBT has some benefits, including dose homogeneity in the target volume as well as dose reduction in deeper tissues [4,11].

The standard technique currently being used at our hospital for treating breast cancer includes opposing tangential beams of photon for the whole chest wall region in addition to an interior photon beam for the supraclavicular lymph nodes. To elevate the dose for PTV (supraclavicular lymph nodes) and reduce the dose for healthy and sensitive surrounding organs, we used a new method. In the newly proposed method, we used two photon-electron combinatory beams (the PA photon beam and AP electron beam) for the supraclavicular region and performed the same process for the chest wall region.

Dose to organs at risk

According to our study, the maximum dose delivered to the spinal cord, head of the humerus bone, and thyroid declined significantly with CPET. The deep penetration of photon (being forward due to high energy) led to less disseminated beams, thus accounting for the lower dose delivered to these organs. The spinal cord is regarded as an organ of the series; therefore, reduction in the maximum dose for this organ is of considerable importance.

Radiotherapy might be accompanied by some side effects for the thyroid gland, including hypothyroidism and thyroid cancer. The mean dose of > 20 Gy for neck and vertebral column can be associated with hypothyroidism and thyroid cancer [12,13]. Recent studies reported the occurrence of hypo- or hyper-thyroidism during the first 3-5 years [12-14]. Consequently, it is noteworthy that the dose delivered to the thyroid was < 20 Gy in the present study.

The dose delivered to the heart and lungs was not significantly different between the two methods. In the current study, the mean V25 (%) values were 4.37% ± 8.36% and 4.41% ± 7.07% for CPET and PT, respectively, confirming the optimized result using the combinatory method.

Considering the studies performed on radiotherapy for breast cancer patients, the risk for coronary artery stenosis and heart failure disorders is higher in patients with left-sided cancer than for those with right-sided malignancies. Heartrelated side effects are more probable after 6-12 months from dose reception. Absorbance of doses > 40 Gy by the mediastinal region is associated with an elevated incidence of heart failure and myocardial disorder. For V25 < 10% (with possibility of 1%), coronary artery stenosis may develop after 10-15 years [15]. In this study, the dose absorbed by the heart was lower than the aforementioned doses, and this result is consistent with the findings of Salem et al. [16].

As shown in Table 2, the V20 (%) values of the lung were 25.23% ± 10.78% and 26.98% ± 10.52% for CPET and PT, respectively. Signs of pulmonary fibrosis might be observed for mean doses of > 10-20 Gy and V20 (%) > 30%, and this was reported to be less with CPET than with PT [15]. Likewise, Salem et al. [16] revealed that the V20 (%) value for the lung needed to be optimized for the photon-electron method. Jabbari et al. [17] reported that the dose for the heart was lower in CPET compared with PT. Furthermore, the dose for the lung was elevated in CPET compared with PT.

Dose to planning treatment volume (supraclavicular nodal)

The mean and maximum doses delivered to the supraclavicular lymph nodes increased significantly by the new therapeutic method. Moreover, tumor volume coverage improved using CPET compared with PT. The P-value for V90 was obtained as 0.031, which is indicative of a significant difference between the two techniques.

The aforementioned results demonstrate that CPET has an enhanced coverage compared with PT. Salem et al. [16] also concluded that the coverage was optimized using CPET compared with PT. Likewise, Jabbari et al. [17] stated that the coverage for the supraclavicular lymph nodes was better using CPET than using PT.

Considering our study findings, parallel and opposing fields might lead to homogeneous distribution of beams from the surface to the tissue depth, and the maximum dose will be optimized in the center of PTV. Regarding the thickness of the supraclavicular tissue in the examined patients (< 4 cm) and the penetration depth of the electron beams (< 5 cm), it seems suitable to use AP electron field for these patients.

In addition, using photons with energy of 15 MV and penetration depth higher than PA could account for reduced absorbed doses in the spinal cord and thyroid. However, com-binatory field techniques have some disadvantages, inclu-ding longer treatment process, high treatment costs, and lack of required hardware facilities (multi leaf collimator shortage and fixation methods) [16,18].

In the current study, the use of combinatory fields was associated with reduced doses delivered to exposed organs volume (the spinal cord, thyroid, and head of the humerus bone), favorable coverage of the target volume, and enhanced dose in the supraclavicular lymph nodes. The dose delivered to the heart and lung was not significantly different between the two therapeutic techniques. Homogenous dose distribution and delivery of the maximum dose to the tumoral tissue by increasing the therapeutic fields have driven the efforts toward using these combinatory techniques.

ACKNOWLEDGMENTS

This article is related to a master thesis with code of 1395.151 from Kashan University of Medical Sciences and a research with the number of 1,730 from Arak University of Medical Sciences.

CONFLICTS OF INTEREST

No potential conflicts of interest were disclosed.

Figures
Fig. 1. An example of dose distribution in the supraclavicular region with the photon technique (A) and combinatory photon-electron technique (B).
Fig. 2. An example of the dose volume histogram of the target tissue and organs at risk calculated for one of the patients treated with combinatory photon-electron technique and photon technique.
Fig. 3. An example of the statistical data table of the target tissue and organs at risk calculated by the ISO-gray software.
Fig. 4. Chart of the maximum delivered doses (Dmax) to the organ at risk and supraclavicular lymph nodes determined by the two techniques. PT, photon technique; CPET, combinatory photon-electron technique.
Fig. 5. The mean delivered doses (Dmean) to the organs at risk and supraclavicular lymph nodes determined by the two techniques. PT, photon technique; CPET, combinatory photon-electron technique.
Fig. 6. The total planning treatment volume V90 (%) and V95 (%) determined by photon technique (PT) and combinatory photon-electron technique (CPET).
Tables

Dosimetric parameters and statistical analysis obtained by one-way analysis of variance test for the two therapeutic methods

Plan Number Mean SD SEM Sig-value
Dmax Cord PT 50 12.55 6.2 1.30 0.014*
CPET 50 8.89 4.68 0.66
Dmax Heart PT 50 47.95 5.96 0.84 0.787
CPET 50 47.62 6.45 0.91
Dmax Lung PT 50 51.2 1.89 0.28 0.629
CPET 50 50.98 2.52 0.36
Dmax Thyroid PT 50 39.26 14.03 1.98 0.042*
CPET 50 34.89 11.54 1.63
Dmax Humerus PT 50 35.58 8.02 1.13 0.004*
CPET 50 31.00 7.25 1.026
Dmax Supra PT 50 51.89 2.71 1.37 0.086
CPET 50 54.30 1.59 0.22
Dmean Cord PT 50 1.18 0.808 0.11 0.111
CPET 50 1.51 1.24 0.18
Dmean Heart PT 50 5.51 3.28 0.46 0.949
CPET 50 5.47 3.29 0.47
Dmean Lung PT 50 12.99 4.24 0.60 0.429
CPET 50 12.33 4.12 0.58
Dmean Thyroid PT 50 6.19 5.55 0.78 0.679
CPET 50 5.78 4.43 0.63
Dmean Humerus PT 50 7.27 5.45 0.77 0.508
CPET 50 7.94 4.77 0.67
Dmean Supra PT 50 41.12 2.98 0.85 0.007*
CPET 50 44.56 1.90 0.27

Sig-value, significant value; PT, photon technique; CPET, combinatory photon-electron technique; Dmax, max dosimetry; Dmean, mean dosimetry. *Results showing significant differences (P < 0.05) of dosimetric parameters between CPET and PT performed for mastectomy patients undergoing supraclavicular nodal irradiation.


Vx (%) parameters for different organs determined by PT and CPET

Plan Number Mean SD SEM Sig-value
Heart V5 (%)a PT 50 32.07 9.75 1.38 0.934
CPET 50 31.41 9.76 1.38
Heart V25 (%)a PT 50 4.41 7.07 1 0.842
CPET 50 4.37 8.36 1.18
Heart V30 (%)a PT 50 3.82 6.81 0.96 0.948
CPET 50 3.71 7.92 1.12
Lung V40 (%)b PT 50 9.43 7.81 1.10 0.797
CPET 50 10.29 10.10 1.67
Lung V20 (%)b PT 50 26.98 10.52 1.49 0.415
CPET 50 25.23 10.78 1.53
Supra V95 (%)c PT 50 55.53 12.31 1.74 0.018*
CPET 50 65.80 13.10 1.85
Supra V90 (%)c PT 50 74.61 9.30 1.31 0.031*
CPET 50 82.06 9.70 1.37

PT, photon technique; CPET, combinatory photon-electron technique; Sig-value, significant value. *Means and statistical analysis results showing significant differences (P < 0.05) of dosimetric parameters between CPET and PT performed for mastectomy patients undergoing supraclavicular nodal irradiation. aThe percentage of the heart volume receiving 5, 25, and 30 Gy dose, respectively. bThe percentage of the ipsilateral lung volume receiving 40 and 20 Gy dose, respectively; cThe percentage of the target tissue volume irradiated to 95 and 90% of the prescribed dose, respectively.


References
  1. DeSantis CE, Fedewa SA, Goding Sauer A, Kramer JL, Smith RA, Jemal A. Breast cancer statistics, 2015: convergence of incidence rates between black and white women. CA Cancer J Clin 2016;66:31-42.
    Pubmed CrossRef
  2. Yang B, Dong Z, Lin MH, Ma CM. A new method to deliver supraclavicular radiation in breast radiotherapy for lung sparing. J Appl Clin Med Phys 2011;12:169-77.
    Pubmed KoreaMed CrossRef
  3. Cakir A, Akgun Z, Fayda M, Agaoglu F. Comparison of three dimensional conformal radiation therapy, intensity modulated radiation therapy and volumetric modulated arc therapy for low radiation exposure of normal tissue in patients with prostate cancer. Asian Pac J Cancer Prev 2015;16:3365-70.
    Pubmed CrossRef
  4. Khan FM. Khan's lectures: handbook of the physics of radiation therapy. Philadelphia, Lippincott Williams & Wilkins, 2010.
  5. Castro Pena P, Kirova YM, Campana F, Dendale R, Bollet MA, Fournier-Bidoz N, et al. Anatomical, clinical and radiological delineation of target volumes in breast cancer radiotherapy planning: individual variability, questions and answers. Br J Radiol 2009;82:595-9.
    Pubmed CrossRef
  6. Beasley M, Driver D, Dobbs HJ. Complications of radiotherapy: improving the therapeutic index. Cancer Imaging 2005;5:78-84.
    Pubmed KoreaMed CrossRef
  7. Remouchamps VM, Vicini FA, Sharpe MB, Kestin LL, Martinez AA, Wong JW. Significant reductions in heart and lung doses using deep inspiration breath hold with active breathing control and intensity-modulated radiation therapy for patients treated with locoregional breast irradiation. Int J Radiat Oncol Biol Phys 2003;55:392-406.
    Pubmed CrossRef
  8. Olofsson L. .
  9. Allen SJ, Klein EE, Michaletz-Lorenz M, Jin JY. Comparison of two treatment techniques for breast irradiation including internal mammary nodes. Med Dosim 2004;29:124-7.
    Pubmed CrossRef
  10. Miles EA, Venables K, Hoskin PJ, Aird EG. Dosimetry and field matching for radiotherapy to the breast and supraclavicular fossa. Radiother Oncol 2009;91:42-8.
    Pubmed CrossRef
  11. Hogstrom KR, Almond PR. Review of electron beam therapy physics. Phys Med Biol 2006;51:R455-89.
    Pubmed CrossRef
  12. Ronckers CM, Sigurdson AJ, Stovall M, Smith SA, Mertens AC, Liu Y, et al. Thyroid cancer in childhood cancer survivors: a detailed evaluation of radiation dose response and its modifiers. Radiat Res 2006;166:618-28.
    Pubmed CrossRef
  13. Sklar C, Whitton J, Mertens A, Stovall M, Green D, Marina N, et al. Abnormalities of the thyroid in survivors of Hodgkin's disease: data from the Childhood Cancer Survivor Study. J Clin Endocrinol Metab 2000;85:3227-32.
    Pubmed CrossRef
  14. Levitt SH, Purdy JA, Perez CA, Poortmans P. Technical basis of radiation therapy: practical clinical applications. Berlin, Springer, 2012.
    CrossRef
  15. Hope A, Karamboulas C, Xu W, Huang S, Kim J, Bratman S, et al. OC-022: Association of patient derived xenograft formation with oral cavity squamous cell cancer outcomes. Radiother Oncol 2017;122(Suppl 1):15.
    CrossRef
  16. Salem A, Mohamad I, Dayyat A, Kanaa'n H, Sarhan N, Roujob I, et al. Combined photon-electron beams in the treatment of the supraclavicular lymph nodes in breast cancer: a novel technique that achieves adequate coverage while reducing lung dose. Med Dosim 2015;40:210-7.
    Pubmed CrossRef
  17. Jabbari K, Azarmahd N, Babazade S, Amouheidari A. Optimizing of the tangential technique and supraclavicular fields in 3 dimensional conformal radiation therapy for breast cancer. J Med Signals Sens 2013;3:107-16.
    Pubmed KoreaMed CrossRef
  18. Tenhunen M, Nyman H, Strengell S, Vaalavirta L. Linac-based isocentric electron-photon treatment of radically operated breast carcinoma with enhanced dose uniformity in the field gap area. Radiother Oncol 2009;93:80-6.
    Pubmed CrossRef